Suspensory Fixation of Grafts in Anterior Cruciate Ligament Reconstruction: A Biomechanical Comparison of 3 Implants

  • Florian S. Kamelger, M.D.

      Affiliations

    • Department of Trauma Surgery and Sports Medicine, Medical University Innsbruck, Innsbruck, Austria
    • Corresponding Author InformationAddress correspondence and reprint requests to Florian S. Kamelger, M.D., Department of Trauma Surgery and Sports Medicine, Medical University Innsbruck, Anichstrasse 35, A-6020 Innsbruck, Austria
  • ,
  • Ursula Onder, M.D.

      Affiliations

    • Department of Trauma Surgery and Sports Medicine, Medical University Innsbruck, Innsbruck, Austria
  • ,
  • Werner Schmoelz, Dr. Dipl.Ing., Ph.D.

      Affiliations

    • Department of Trauma Surgery and Sports Medicine, Medical University Innsbruck, Innsbruck, Austria
  • ,
  • Katja Tecklenburg, M.D.

      Affiliations

    • Department of Trauma Surgery and Sports Medicine, Medical University Innsbruck, Innsbruck, Austria
  • ,
  • Rohit Arora, M.D.

      Affiliations

    • Department of Trauma Surgery and Sports Medicine, Medical University Innsbruck, Innsbruck, Austria
  • ,
  • Christian Fink, M.D.

      Affiliations

    • Department of Trauma Surgery and Sports Medicine, Medical University Innsbruck, Innsbruck, Austria
    • Sportsclinic Austria, Innsbruck, Austria

Received 12 September 2008; accepted 21 January 2009. published online 27 April 2009.

Purpose

Our purpose was to evaluate the biomechanical difference between 2 new soft-tissue anterior cruciate ligament (ACL) fixation devices (RetroButton [Arthrex, Naples, FL] and ToggleLoc [Biomet, Warsaw, IN]) and a clinically established implant (EndoButton CL; Smith & Nephew, Andover, MA).

Methods

In test session 1 biomechanical testing was performed by use of 48 porcine femora fixed to a material testing machine. In session 2 the implants were tested alone. Cyclic loading was carried out for 1,000 cycles, followed by a load-to-failure test. The tested devices were as follows: EndoButton CL, 20-mm and 40-mm loops; RetroButton, 12/20-mm and 12/40-mm loops; and ToggleLoc, 20-mm and 40-mm loops.

Results

In test session 1 the EndoButton CL showed higher motion per cycle (0.30 ± 0.09 mm for 20-mm loop and 0.43 ± 0.03 mm for 40-mm loop) but lower plastic displacement values (0.66 ± 0.19 mm for 20-mm loop and 0.50 ± 0.13 mm for 40-mm loop) than the objects of comparison. The RetroButton devices were the stiffest (331.47 ± 133.92 N/mm for 12/20-mm loop and 265.66 ± 103.97 N/mm for 12/40-mm loop). The highest ultimate failure load, however, was shown by the EndoButton CL devices. In test session 2 the EndoButton CL oscillated the most. The ToggleLoc devices elongated more than the objects of comparison (0.66 ± 0.12 mm for 20-mm loop and 0.76 ± 0.06 mm for 40-mm loop). The RetroButton devices were both the stiffest among the respective implants (542.7 ± 148.0 N/mm for 12/20-mm loop and 379.0 ± 40.1 N/mm for 12/40-mm loop). The EndoButton CL showed the highest values for displacement to failure (3.6 ± 0.4 mm for 20-mm loop and 6.4 ± 0.4 mm for 40-mm loop). The ToggleLoc devices failed the latest on load-to-failure testing.

Conclusions

All tested implants could provide adequate fixation strength. Despite advantages in the design of the EndoButton CL regarding its handling, the 2 newly released products showed superior material properties.

Clinical Relevance

Suspensory fixation of hamstring grafts in ACL reconstruction is frequently associated with bone tunnel enlargement. Material properties and implant design may limit graft-tunnel motion and result in enlargement of the femoral bone tunnel. With improved implant design, suspensory graft fixation may still be an attractive fixation technique in primary and revision ACL reconstruction.

Key Words: Anterior cruciate ligament reconstruction, Suspensory fixation, Knee, EndoButton CL, RetroButton, ToggleLoc

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 The authors report no conflict of interest.

PII: S0749-8063(09)00075-9

doi:10.1016/j.arthro.2009.01.021

Arthroscopy: The Journal of Arthroscopic and Related Surgery
Volume 25, Issue 7 , Pages 767-776, July 2009