Thermal profile of radiofrequency energy in the inferior glenohumeral ligament

      Purpose

      Currently, two different methods of applying radiofrequency (RF) energy (monopolar and bipolar) are available to the surgeon for thermal shortening of the shoulder capsule. The objective of this study was to investigate the temperature changes and the thermal conduction across the human inferior glenohumeral ligament (IGHL) during radiofrequency energy application.

      Methods

      Thermistors were secured onto both the intra-articular and extra-articular surfaces of human IGHL. Monopolar RF energy and bipolar RF energy were delivered to the intra-articular surface at the manufacturer’s recommended settings. Pre-treatment and post-treatment ligament lengths, widths, heating times, and temperatures were measured and compared.

      Results

      For the monopolar devices, temperature spikes to 89°C were recorded for the set temperature of 67°C, averaging 77°C ± 10°C. Temperatures across the ligament averaged 48°C ± 3°C. For both devices, the IGHL became thicker with higher RF settings. Recorded temperatures decreased as distance increased from the point of application. Maximum temperatures occurred at least 6 to 7 seconds after cessation of energy application.

      Conclusions

      The bipolar and monopolar devices had similar conduction times across the ligament, suggesting that this occurs by simple diffusion of heat. Bipolar and monopolar devices were equally efficacious for capsular shrinkage if the extent of the shortening is tightly defined.

      Clinical relevance

      The thermal probe should not rest in one position for an extended period of time during RF energy application because, as our study showed, the monitoring of temperature or the visualization of tissue change is not efficacious for determining the end point of thermal shrinkage of the shoulder capsule.

      Key words

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Arthroscopy
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Blouin L.T.
        • Marcus F.I.
        • Lampe L.
        Assessment of effects of a radiofrequency field and temperature location in an electrode catheter on the accuracy of temperature measurement.
        PACE. 1991; 14: 807-813
        • Dinerman J.L.
        • Berger R.D.
        • Calkins H.
        Temperature monitoring during radiofrequency ablation.
        J Cardiovasc Electrophysiol. 1996; 7: 161-173
        • Goldberg S.N.
        • Gazelle G.S.
        • Dawson S.L
        • et al.
        Tissue ablation with radiofrequency.
        Acad Radiol. 1995; 2: 399-404
        • Otomo K.
        • Yamanashi W.S.
        • Tondo C
        • et al.
        Why a large tip electrode makes a deeper radiofrequency lesion.
        J Cardiovasc Electrophysiol. 1998; 9: 47-54
        • Levy O.
        • Wilson M.
        • Williams H
        • et al.
        Thermal capsular shrinkage for shoulder instability mid term longitudinal outcome study.
        J Bone Joint Surg Br. 2001; 83: 640-645
        • Mishra D.K.
        • Fanton G.S.
        Two-year outcome of arthroscopic Bankart repair and electrothermal-assisted capsulorrhaphy for recurrent traumatic anterior shoulder instability.
        Arthroscopy. 2001; 17: 844-849
        • Hayashi K.
        • Thabit G.
        • Massa K.L
        • et al.
        The effect of thermal heating on the length and histological properties of the glenohumeral joint capsule.
        Am J Sports Med. 1997; 25: 107-112
        • Naseef G.S.
        • Foster T.E.
        • Trauner K
        • et al.
        The thermal properties of the bovine joint capsule.
        Am J Sports Med. 1997; 25: 670-674
        • Vangsness Jr, C.T.
        • Mitchell W.
        • Nimni M
        • et al.
        Collagen shortening.
        Clin Orthop Rel Res. 1997; 337: 267-271
        • Wallace A.L.
        • Hollinshead R.M.
        • Frank C.B.
        Creep behavior of a rabbit model of ligament laxity after electrothermal shrinkage in vivo.
        Am J Sports Med. 2002; 30: 98-9102
        • Gryler E.C.
        • Greis P.E.
        • Burks R.T.
        • West J.
        Axillary nerve temperatures drug radiofrequency capsulorrhaphy of the shoulder.
        Arthroscopy. 2001; 17: 567-572
        • Haines D.E.
        • Watson D.D.
        • Verow A.F.
        Electrode radius predicts lesion radius during radiofrequency energy heating.
        Circ Res. 1990; 67: 124-129
        • Hayashi K.
        • Markel M.D.
        Thermal capsulorrhaphy treatment of shoulder instability.
        Clin Orthop. 2001; 390: 59-72
        • Foster T.E.
        • Elman M.
        Arthroscopic delivery systems used for thermally induced shoulder capsulorrhaphy.
        Oper Tech Sports Med. 1998; 6: 126-130
        • Wong K.L.
        • Williams G.R.
        Complications of thermal capsulorrhaphy of the shoulder.
        J Bone Joint Surg Am. 2001; 83: 151-155
        • Greis P.E.
        • Burks R.T.
        • Schinkendantz M.S.
        • Sandmeier R.
        Axillary nerve injury after thermal capsular shrinkage of the shoulder.
        J Shoulder Elbow Surg. 2001; 10: 231-235
        • Nath S.
        • DiMarco J.P.
        • Haines D.E.
        Basic aspects of radiofrequency catheter ablation.
        J Cardiovasc Electrophysiol. 1994; 5: 863-876
        • Shellock F.G.
        • Shields Jr, C.L.
        Temperature changes associated with radiofrequency energy-induced heating of bovine capsular tissue.
        Am J Sports Med. 2000; 16: 348-358
        • Obrzut S.L.
        • Hecht P.
        • Hayashi K
        • et al.
        The effect of radiofrequency energy on the length and temperature properties of the glenohumeral joint capsule.
        Arthroscopy. 1998; 14: 395-400