Two Intramedullary and One Extramedullary Cortical Button, With or Without Interference Screw, Show Biomechanical Properties Superior to Native Tendon in Repair of the Distal Biceps Tendon: A Systematic Review and Network Meta-analysis of Biomechanical Performance

Published:October 12, 2022DOI:https://doi.org/10.1016/j.arthro.2022.08.037

      Abstract

      Purpose

      The aim of this study was to use a systematic review and network meta-analysis (NMA) to compare the failure strength, maximum strength, stiffness, and displacement of available constructs for distal biceps repair.

      Methods

      An NMA was conducted to determine the performance of 2 all-suture suture anchors (2x ASA), 2 intramedullary cortical buttons (2x IM CB), 2 suture anchors (2x SA), extramedullary cortical buttons (EM CB), extramedullary cortical button plus interference screw (EM CB+IFS), interference screw (IFS), single intramedullary cortical button (IM CB), single suture anchor (SA), trans-osseous suture (TOS), tension slide technique (TST), and tension slide technique plus suture tape (TST+ST). Analysis consisted of arm-based network meta-analysis under Bayesian random-effects model with Markov Chain Monte Carlo (MCMC) sampling. Biomechanical outcomes were summarized as treatment effects and their corresponding 95% confidence intervals (CI). Rank probabilities were calculated and used to generate each treatment’s surface under the cumulative ranking (SUCRA) curve. Biomechanical properties were compared to native tendon. Displacement >10mm was defined as clinical failure.

      Results

      Twenty-one studies were included. For failure strength, no construct outperformed the native tendon but 2x SA, IFS, SA, and TOS demonstrated poorer failure strength. For the maximum load to failure, EM CB+IFS outperformed the native tendon. Compared to native tendon, EM CB+IFS, EM CB, and 2xIM CB were stiffer, while 2x SA and IFS were less stiff. No construct demonstrated >10 mm of displacement, but constructs with displacement above the mean (3.5 mm) included 2x ASA, 2xIM CB, and TOS.

      Conclusions

      The fixation constructs that consistently demonstrated comparable or better biomechanical properties (failure strength, maximum strength, stiffness) to native tendon in distal biceps tendon repair were the extramedullary cortical button with or without interference screw and two intramedullary cortical buttons. No construct demonstrated displacement beyond standard definitions for clinical failure.

      Clinical Relevance

      This network meta-analysis of biomechanical studies suggest that extramedullary cortical button and two intramedullary cortical buttons may be the most stable construct for distal biceps repair fixation, with equivalent or better biomechanical properties compared to native tendon.

      Key terms

      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Arthroscopy
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      REFERENCES

        • D'Alessandro D.F.
        • Shields Jr., C.L.
        • Tibone J.E.
        • Chandler R.W.
        Repair of distal biceps tendon ruptures in athletes.
        Am J Sports Med. 1993; 21: 114-119
        • Kelly M.P.
        • Perkinson S.G.
        • Ablove R.H.
        • Tueting J.L.
        Distal Biceps Tendon Ruptures: An Epidemiological Analysis Using a Large Population Database.
        Am J Sports Med. 2015; 43: 2012-2017
        • Safran M.R.
        • Graham S.M.
        Distal biceps tendon ruptures: incidence, demographics, and the effect of smoking.
        Clin Orthop Relat Res. 2002; : 275-283
        • Morrey B.F.
        • Askew L.J.
        • An K.N.
        • Dobyns J.H.
        Rupture of the distal tendon of the biceps brachii. A biomechanical study.
        J Bone Joint Surg Am. 1985; 67: 418-421
        • Athwal G.S.
        • Steinmann S.P.
        • Rispoli D.M.
        The distal biceps tendon: footprint and relevant clinical anatomy.
        J Hand Surg Am. 2007; 32: 1225-1229
        • Siebenlist S.
        • Lenich A.
        • Buchholz A.
        • et al.
        Biomechanical in vitro validation of intramedullary cortical button fixation for distal biceps tendon repair: a new technique.
        Am J Sports Med. 2011; 39: 1762-1768
        • Baker B.E.
        • Bierwagen D.
        Rupture of the distal tendon of the biceps brachii. Operative versus non-operative treatment.
        J Bone Joint Surg Am. 1985; 67: 414-417
        • Lynch S.A.
        • Beard D.M.
        • Renstrom P.A.
        Repair of distal biceps tendon rupture with suture anchors.
        Knee Surg Sports Traumatol Arthrosc. 1999; 7: 125-131
        • Berlet G.C.
        • Johnson J.A.
        • Milne A.D.
        • Patterson S.D.
        • King G.J.
        Distal biceps brachii tendon repair. An in vitro biomechanical study of tendon reattachment.
        Am J Sports Med. 1998; 26: 428-432
        • Balabaud L.
        • Ruiz C.
        • Nonnenmacher J.
        • Seynaeve P.
        • Kehr P.
        • Rapp E.
        Repair of distal biceps tendon ruptures using a suture anchor and an anterior approach.
        J Hand Surg Br. 2004; 29: 178-182
        • Idler C.S.
        • Montgomery 3rd, W.H.
        • Lindsey D.P.
        • Badua P.A.
        • Wynne G.F.
        • Yerby S.A.
        Distal biceps tendon repair: a biomechanical comparison of intact tendon and 2 repair techniques.
        Am J Sports Med. 2006; 34: 968-974
        • Moher D.
        • Liberati A.
        • Tetzlaff J.
        • Altman D.G.
        • Group P.
        Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement.
        Ann Intern Med. 2009; 151 (W264): 264-269
        • Sterne J.A.C.
        • Hernán M.A.
        • Reeves B.C.
        • et al.
        ROBINS-I: a tool for assessing risk of bias in non-randomised studies of interventions.
        British Medical Journal. 2016;
        • Cohen W.R.
        • Ommani S.
        • Hassan S.
        • et al.
        Accuracy and reliability of fetal heart rate monitoring using maternal abdominal surface electrodes.
        Acta Obstet Gynecol Scand. 2012; 91: 1306-1313
        • Tonin F.S.
        • Rotta I.
        • Mendes A.M.
        • Pontarolo R.
        Network meta-analysis: a technique to gather evidence from direct and indirect comparisons.
        Pharm Pract (Granada). 2017; 15: 943
        • Lin L.
        • Zhang J.
        • Hodges J.S.
        • Chu H.
        Performing Arm-Based Network Meta-Analysis in R with the pcnetmeta Package.
        Journal of statistical software. 2017; 80
        • Lu G.
        • Ades A.E.
        Combination of direct and indirect evidence in mixed treatment comparisons.
        Stat Med. 2004; 23: 3105-3124
        • Higgins J.P.
        • Jackson D.
        • Barrett J.K.
        • Lu G.
        • Ades A.E.
        • White I.R.
        Consistency and inconsistency in network meta-analysis: concepts and models for multi-arm studies.
        Res Synth Methods. 2012; 3: 98-110
        • Efthimiou O.
        • Debray T.P.
        • van Valkenhoef G.
        • et al.
        GetReal in network meta-analysis: a review of the methodology.
        Res Synth Methods. 2016; 7: 236-263
        • Pereira D.S.
        • Kvitne R.S.
        • Liang M.
        • Giacobetti F.B.
        • Ebramzadeh E.
        Surgical repair of distal biceps tendon ruptures: a biomechanical comparison of two techniques.
        Am J Sports Med. 2002; 30: 432-436
        • Spang J.T.
        • Weinhold P.S.
        • Karas S.G.
        A biomechanical comparison of EndoButton versus suture anchor repair of distal biceps tendon injuries.
        J Shoulder Elbow Surg. 2006; 15: 509-514
        • Failla J.M.
        • Amadio P.C.
        • Morrey B.F.
        • Beckenbaugh R.D.
        Proximal radioulnar synostosis after repair of distal biceps brachii rupture by the two-incision technique. Report of four cases.
        Clin Orthop Relat Res. 1990; : 133-136
        • Chavan P.R.
        • Duquin T.R.
        • Bisson L.J.
        Repair of the ruptured distal biceps tendon: a systematic review.
        Am J Sports Med. 2008; 36: 1618-1624
        • Miyamoto R.G.
        • Elser F.
        • Millett P.J.
        Distal biceps tendon injuries.
        J Bone Joint Surg Am. 2010; 92: 2128-2138
        • Agrawal V.
        • Stinson M.J.
        Case report: heterotopic ossification after repair of distal biceps tendon rupture utilizing a single-incision Endobutton technique.
        J Shoulder Elbow Surg. 2005; 14: 107-109
        • Bain G.I.
        • Prem H.
        • Heptinstall R.J.
        • Verhellen R.
        • Paix D.
        Repair of distal biceps tendon rupture: a new technique using the Endobutton.
        J Shoulder Elbow Surg. 2000; 9: 120-126
        • Peeters T.
        • Ching-Soon N.G.
        • Jansen N.
        • Sneyers C.
        • Declercq G.
        • Verstreken F.
        Functional outcome after repair of distal biceps tendon ruptures using the endobutton technique.
        J Shoulder Elbow Surg. 2009; 18: 283-287
        • Arianjam A.
        • Camisa W.
        • Leasure J.M.
        • Montgomery W.H.
        Biomechanical comparison of interference screw and cortical button with screw hybrid technique for distal biceps brachii tendon repair.
        Orthopedics. 2013; 36: e1371-1377
        • Sethi P.
        • Obopilwe E.
        • Rincon L.
        • Miller S.
        • Mazzocca A.
        Biomechanical evaluation of distal biceps reconstruction with cortical button and interference screw fixation.
        J Shoulder Elbow Surg. 2010; 19: 53-57
        • Rose D.M.
        • Archibald J.D.
        • Sutter E.G.
        • Belkoff S.M.
        • Wilckens J.H.
        Biomechanical analysis suggests early rehabilitation is possible after single-incision EndoButton distal biceps repair with FiberWire.
        Knee Surg Sports Traumatol Arthrosc. 2011; 19: 1019-1022
        • Lemos S.E.
        • Ebramzedeh E.
        • Kvitne R.S.
        A new technique: in vitro suture anchor fixation has superior yield strength to bone tunnel fixation for distal biceps tendon repair.
        Am J Sports Med. 2004; 32: 406-410
        • Klein D.M.
        • Ghany N.
        • Urban W.
        • Caruso S.A.
        Distal biceps tendon repair: anchor versus transosseous suture fixation.
        Am J Orthop (Belle Mead NJ). 2007; 36: 34-37
        • Kettler M.
        • Lunger J.
        • Kuhn V.
        • Mutschler W.
        • Tingart M.J.
        Failure strengths in distal biceps tendon repair.
        Am J Sports Med. 2007; 35: 1544-1548
        • Greenberg J.A.
        • Fernandez J.J.
        • Wang T.
        • Turner C.
        EndoButton-assisted repair of distal biceps tendon ruptures.
        J Shoulder Elbow Surg. 2003; 12: 484-490
        • Siebenlist S.
        • Buchholz A.
        • Zapf J.
        • et al.
        Double intramedullary cortical button versus suture anchors for distal biceps tendon repair: a biomechanical comparison.
        Knee Surg Sports Traumatol Arthrosc. 2015; 23: 926-933
        • Mazzocca A.D.
        • Burton K.J.
        • Romeo A.A.
        • Santangelo S.
        • Adams D.A.
        • Arciero R.A.
        Biomechanical evaluation of 4 techniques of distal biceps brachii tendon repair.
        Am J Sports Med. 2007; 35: 252-258
        • Majumdar A.
        • Chavez W.
        • Valdez J.
        • et al.
        Unicortical Versus Bicortical Button Fixation For Distal Biceps Brachii Tendon Rupture: A Cadaveric Biomechanical Study.
        Orthopaedic Journal of Sports Medicine. 2019; 7 (supp 5): 7
        • An K.N.
        • Hui F.C.
        • Morrey B.F.
        • Linscheid R.L.
        • Chao E.Y.
        Muscles across the elbow joint: a biomechanical analysis.
        J Biomech. 1981; 14: 659-669
        • Kodde I.F.
        • Baerveldt R.C.
        • Mulder P.G.
        • Eygendaal D.
        • van den Bekerom M.P.
        Refixation techniques and approaches for distal biceps tendon ruptures: a systematic review of clinical studies.
        J Shoulder Elbow Surg. 2016; 25: e29-37
        • Siebenlist S.
        • Schmitt A.
        • Imhoff A.B.
        • et al.
        Intramedullary Cortical Button Repair for Distal Biceps Tendon Rupture: A Single-Center Experience.
        J Hand Surg Am. 2019; 44: 418 e411-418 e417